Lutron PH-222

Akwụkwọ ntuziaka onye ọrụ Lutron PH-222 Elektrọniki Pen pH

1. Okwu mmalite

This manual provides comprehensive instructions for the proper setup, operation, calibration, and maintenance of the Lutron PH-222 Electronic Pen pH Meter. The PH-222 is a portable, high-precision instrument designed for accurate pH measurements in various applications, including water quality testing, laboratory analysis, and industrial processes. Adherence to these instructions will ensure optimal performance and longevity of the device.

2. Ozi nchekwa

Please read and understand all safety instructions before operating the Lutron PH-222 pH Meter. Failure to follow these guidelines may result in injury or damage to the instrument.

  • Always handle the pH electrode with care. The glass bulb is fragile.
  • Do not ingest pH buffer solutions. In case of accidental ingestion, seek medical attention immediately.
  • Debe ngwaọrụ ahụ ka ọ ghara inwe oke okpomọkụ, anwụ kpọmkwem, na oke iru mmiri.
  • Ensure the battery compartment is properly sealed to maintain the IP-67 waterproof rating.
  • Tụfuo batrị eji eme ihe dịka iwu mpaghara siri dị.
  • Do not attempt to disassemble or modify the meter, as this will void the warranty and may cause damage.

3. Ihe ngwugwu

Mgbe ebupuchara, chọpụta na ihe niile edepụtara n'okpuru dị ugbu a yana enweghị mmebi:

  • Lutron PH-222 Electronic Pen pH Meter
  • Akwụkwọ ntuziaka ọrụ
  • Asambodo mmezi
  • pH 4.00 Buffer Solution (small bottle)
  • pH 7.00 Buffer Solution (small bottle)
Lutron PH-222 Pen pH Meter and accessories in packaging

Foto 3.1: Contents of the Lutron PH-222 package. This image displays the PH-222 pen pH meter, its operation manual, a calibration certificate, and two small bottles of pH buffer solutions (pH 4.00 and pH 7.00).

4. Akụrụngwa Ngwaahịa

The Lutron PH-222 pH Meter incorporates several features designed for user convenience and accurate measurement:

  • Easy to Change the pH Electrode: Facilitates quick replacement of the sensing electrode.
  • High Input Impedance: Ensures minimal current draw from the pH electrode, leading to stable and accurate readings.
  • Data Hold Function: Allows freezing of the displayed value for convenient recording.
  • Auto Power Off: Conserves battery life by automatically shutting down after a period of inactivity.
  • 100 Point Data Logger with Recall: Stores up to 100 measurement points for later review.
  • IP-67 Waterproof Rating: Provides protection against dust and immersion in water up to 1 meter for 30 minutes.
N'ihu view of Lutron PH-222 Electronic Pen pH Meter

Foto 4.1: N'ihu view of the Lutron PH-222 Electronic Pen pH Meter, showing the display and control buttons (POWER, HOLD, REC).

5. Nhazi

5.1 Ntinye batrị

The Lutron PH-222 requires batteries for operation (not included in the package). To install or replace batteries:

  1. Twist the top cap of the meter counter-clockwise to open the battery compartment.
  2. Insert two 1.5V AAA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
  3. Securely close the battery compartment by twisting the cap clockwise until it is tight to maintain the waterproof seal.

5.2 Electrode Preparation

  1. Carefully remove the protective cap from the bottom of the pH electrode. The cap typically contains a small amount of storage solution to keep the electrode moist.
  2. Rinse the electrode tip with distilled or deionized water to remove any residual storage solution. Gently blot dry with a lint-free tissue. Do not rub the glass bulb.

6. Ọrụ

6.1 Gbanyụọ ma ọ bụ gbanyụọ

Pịa IKE bọtịnụ iji gbanye ma ọ bụ gbanyụọ mita ahụ.

6.2 Inyocha ihe

  1. Ensure the meter is calibrated (refer to Section 7).
  2. Immerse the pH electrode tip into the sample solution, ensuring the glass bulb is fully submerged.
  3. Gently stir the solution to ensure homogeneity and allow the reading to stabilize. The display will show the pH value.
  4. Record the stable pH reading.
  5. After measurement, rinse the electrode with distilled water and replace the protective cap.

6.3 Data Jide Ọrụ

Iji mee ka ihe dị ugbu a dị na ihuenyo ahụ gbụchie, pịa bọtịnụ Jide button. Press it again to release the hold and return to live measurement.

6.4 Data Logging and Recall

The PH-222 can store up to 100 measurement points. Refer to the detailed operation manual included in the package for specific instructions on how to use the REC button for data logging and recall functions.

7. nhazi nhazi

Regular calibration is essential for accurate pH measurements. The Lutron PH-222 supports multi-point calibration using standard buffer solutions. It is recommended to calibrate the meter before each use or if readings appear inaccurate.

7.1 Calibration Procedure (Two-Point)

The PH-222 typically uses a two-point calibration with pH 7.00 and pH 4.00 (or pH 10.00) buffer solutions.

  1. Prepare fresh pH 7.00 and pH 4.00 buffer solutions. Ensure they are at the same temperature as your samples.
  2. Turn on the meter and rinse the electrode with distilled water.
  3. Immerse the electrode into the pH 7.00 buffer solution.
  4. Pịa ma jide CAL button (often integrated with the REC button, refer to the device's specific button functions) until the calibration mode is activated. The display will typically show 'CAL' or a flashing pH value.
  5. Wait for the reading to stabilize. The meter will automatically recognize the buffer and complete the calibration for pH 7.00.
  6. Jiri mmiri distilled sachaa electrode ahụ.
  7. Immerse the electrode into the pH 4.00 buffer solution.
  8. Repeat the calibration process. The meter will automatically recognize the buffer and complete the calibration for pH 4.00.
  9. Once calibration is complete, the meter will return to measurement mode.
Bottles of pH 4.00 and pH 7.00 buffer solutions

Foto 7.1: Standard pH 4.00 and pH 7.00 buffer solutions used for calibrating the pH meter.

8. Nlekọta

8.1 Electrode Care and Storage

  • Always keep the electrode tip moist. Never allow the electrode to dry out. Store it in its protective cap containing a small amount of storage solution (or 3M KCl solution).
  • Rinse the electrode with distilled water after each use to prevent contamination.
  • If the electrode becomes dirty or fouled, clean it gently according to the instructions in the full operation manual.
  • Replace the electrode when it no longer calibrates correctly or provides stable readings.

8.2 Nhicha izugbe

Wipe the meter body with a damp akwa . Ejila ihe nchacha abrasive ma ọ bụ ihe mgbaze.

8.3 nnọchi batrị

Replace batteries when the low battery indicator appears on the display or if the meter fails to power on. Refer to Section 5.1 for battery installation instructions.

9. Nchọpụta nsogbu

NsogbuIhe nwere ike imeNgwọta
Mita anaghị arụ ọrụBatrị nwụrụ anwụ ma ọ bụ arụnyere na ezighi eziDochie batrị, hụ na ọ dị mma iji ya.
Unstable or erratic readingsDirty electrode; air bubbles on electrode; electrode drying out; old buffer solutionsClean electrode; gently tap to remove bubbles; re-moisten electrode; use fresh buffer solutions.
Ọgụgụ na-ezighi eziMeter not calibrated; electrode nearing end of life; incorrect buffer solutionsPerform calibration (Section 7); replace electrode; use correct and fresh buffer solutions.
Calibration failsDirty or damaged electrode; incorrect buffer solutions; electrode dryClean/replace electrode; ensure correct and fresh buffer solutions are used; ensure electrode is moist.

10. Nkọwapụta

OkeNkọwapụta
Nọmba nlereanyaPH-222
ỤdịLutron
Oke nha0 ka 14.00 pH
Ntinye NtinyeElu
Data Logger100 points with recall
Ịnye ọkụ3 Volts (e.g., 2 x 1.5V AAA batteries, not included)
Gbanyụọ ike akpaakaEe
Ngosi nke enweghị mmiriIP-67
Ibu Ibu300 g
Akụkụ ngwaahịa (LxWxH)18 x 4 x 3.2 cm
Mba onye siTaiwan

11. Akwụkwọ ikike na nkwado

For warranty information, technical support, or service inquiries, please refer to the warranty card included with your product or contact your authorized Lutron dealer or the manufacturer directly. Ensure you have your product model number (PH-222) and purchase details available when seeking support.

Akwụkwọ ndị emetụtara - PH-222

Tupuview Lutron PH-207 pH Meter Operation Manual
This operation manual provides detailed instructions for the Lutron PH-207 pH meter, covering its features, specifications, calibration procedures, measurement techniques, temperature compensation, RS232 interface, battery replacement, and optional accessories.
Tupuview Lutron TU-2016 Turbidity Meter Operation Manual
Comprehensive operation manual for the Lutron TU-2016 Turbidity Meter, detailing features, specifications, measurement procedures, calibration, and battery replacement.
Tupuview Lutron LX-1108 Digital Lux Meter: Nkọwapụta, Ọrụ na atụmatụ
Ntuziaka zuru oke na Lutron LX-1108 dijitalụ lux mita nke Seeanco. Nkọwa nkọwa, usoro arụmọrụ, ndekọ data, na kpachara anya maka nha ọkụ ọkụ ziri ezi.
Tupuview Lutron MO-2014 Micro-Ohm Meter Operation Manual
This document provides detailed operation instructions for the Lutron MO-2014 Micro-Ohm Meter, covering features, specifications, measurement procedures, and troubleshooting.
Tupuview Lutron PWH-303 Pure Water / Hardness Meter Operation Manual
Operation manual for the Lutron PWH-303 pen-type Pure Water/Hardness Meter. Details features, specifications, front panel descriptions, measuring procedures, calibration, battery replacement, and optional accessories.
Tupuview Lutron VB-8206SD Vibration Meter Operation Manual with SD Card Datalogger
Comprehensive operation manual for the Lutron VB-8206SD Vibration Meter, detailing features, specifications, measurement procedures, and SD card data logging capabilities for industrial vibration monitoring.