1. Okwu mmalite
Thank you for choosing the Solight V15 II CAT III Multimeter. This device is designed for safe and accurate measurement of DC/AC voltage, DC current, resistance, diode testing, and continuity. Please read this manual thoroughly before use to ensure proper operation and to prevent damage to the meter or injury to yourself.
Nchekwa Mbụ: Always observe all safety precautions outlined in this manual. Electrical measurements can be dangerous if not performed correctly.
2. Ngwaahịa gafereview

Foto 1: Solight V15 II Multimeter and Test Leads. The image displays the front of the Solight V15 II multimeter, which is dark gray with a large digital display at the top. Below the display is a central rotary switch for selecting measurement functions and ranges. To the left of the multimeter are two test leads: one red and one black, each with a pointed probe tip and a comfortable grip. The multimeter features input jacks at the bottom for connecting these leads.
2.1 Akụkụ
- Ngosipụta Digital: Shows measurement readings, units, and other indicators (e.g., "HV" for high voltage, polarity).
- Ngbanwe Rotary: Used to select the desired measurement function (e.g., V~, V--, A--, Ω, Diode, Continuity) and range.
- Ntinye Jacks:
- "COM" Jack: Ntinye nkịtị (na-adịghị mma) maka nha niile.
- "VΩmA" Jack: Ntinye dị mma maka voltage, resistance, diode, continuity, and current up to 500mA.
- "10A" Jack: Positive input for high current measurements (up to 10A).
- Nleba ule: Red (positive) and Black (common/negative) leads with probe tips for connecting to circuits.
- Akụkụ batrị: Located on the back for 2x AAA batteries.
3. Nhazi
3.1 Ntinye batrị
- Hụ na agbanyụrụla multimeter ahụ.
- Chọta ebe mkpuchi batrị n'azụ unit ahụ.
- Mepee skru(s) ndị ahụ ma wepụ mkpuchi ahụ.
- Insert two (2) AAA (LR03) batteries, observing the correct polarity (+ and -) as indicated inside the compartment.
- Dochie mkpuchi batrị ahụ ma jiri skru(s) mechie ya.
- If the low battery indicator appears on the display, replace the batteries promptly.
3.2 Njikọ Ndị Nnwale Nlele
- Always connect the black test lead to the "COM" jack.
- Maka voltage, resistance, diode, continuity, and current up to 500mA measurements, connect the red test lead to the "VΩmA" jack.
- For current measurements between 500mA and 10A, connect the red test lead to the "10A" jack.
- Ensure connections are firm and secure before taking any measurements.
4. Ntuziaka ọrụ
Before taking any measurement, ensure the test leads are correctly connected and the rotary switch is set to the appropriate function and range. Always start with a higher range if the expected value is unknown to prevent overloading the meter.
4.1 DC Voltage Measurement (V--)
- Tọọ mgba ọkụ rotary na DC Vol achọrọtage (V--) range (e.g., 500V, 200V, 20V, 2000mV, 200mV).
- Connect the black test lead to the "COM" jack and the red test lead to the "VΩmA" jack.
- Connect the test probes across the component or circuit to be measured, observing polarity.
- Gụọ voltage value on the digital display. If a negative sign appears, it indicates reversed polarity.
4.2 AC Voltage Measurement (V~)
- Tọọ mgba ọkụ na-agbanwe agbanwe na AC Vol achọrọtagoke e (V~) (dịka ọmụmaatụ, 500V, 200V).
- Connect the black test lead to the "COM" jack and the red test lead to the "VΩmA" jack.
- Connect the test probes across the AC voltage isi iyi.
- Gụọ voltage uru na dijitalụ ngosi.
4.3 DC Current Measurement (A--)
- Mkpa: Current measurements require the meter to be connected in series with the circuit. Ensure the circuit is de-energized before connecting the meter.
- Set the rotary switch to the desired DC Current (A--) range (e.g., 10A, 200mA, 20mA, 2000µA, 200µA).
- For ranges up to 200mA, connect the red test lead to the "VΩmA" jack. For the 10A range, connect the red test lead to the "10A" jack. The black lead always connects to "COM".
- Open the circuit where current is to be measured and insert the meter in series.
- Apply power to the circuit and read the current value on the display.
- After measurement, disconnect power, remove the meter, and restore the circuit.
4.4 Resistance Measurement (Ω)
- Ịdọ aka ná ntị: Ensure the circuit or component under test is completely de-energized before measuring resistance.
- Set the rotary switch to the desired Resistance (Ω) range (e.g., 20000kΩ, 2000kΩ, 200kΩ, 20kΩ, 2000Ω, 200Ω).
- Connect the black test lead to "COM" and the red test lead to "VΩmA".
- Jikọọ nyocha ule n'ofe akụkụ ahụ.
- Gụọ uru ndị na-eguzogide na ngosipụta.
4.5 Diode Test (→|→)
- Set the rotary switch to the Diode Test position.
- Connect the black test lead to "COM" and the red test lead to "VΩmA".
- Connect the red probe to the anode and the black probe to the cathode of the diode. The display should show a forward voltage drop (typically 0.5V to 0.8V for silicon diodes).
- Reverse the probes. The display should show "OL" (Overload) for a good diode. If it shows a reading in both directions or "OL" in both directions, the diode is likely faulty.
4.6 Continuity Test (•))))
- Set the rotary switch to the Continuity Test position.
- Connect the black test lead to "COM" and the red test lead to "VΩmA".
- Jikọọ ihe nyocha ule n'ofe sekit ma ọ bụ akụkụ ahụ.
- If the resistance is below approximately 50Ω (this value can vary), the buzzer will sound, indicating continuity. The display will also show the resistance value.
5. Nkọwapụta
| Ọrụ nha | Oke | Izi ezi |
|---|---|---|
| DC Voltage (V--) | 200mV, 2000mV, 20V, 200V, 500V | ± 1.0% nke ịgụ + 2 mkpụrụ |
| AC Voltage (V~) | 200V, 500V | ± 1.2% nke ịgụ + 10 mkpụrụ |
| DC Current (A--) | 200µA, 2000µA, 20mA, 200mA, 10A | ± 1.5% nke ịgụ + 2 mkpụrụ |
| Nguzogide (Ω) | 200Ω, 2000Ω, 20kΩ, 200kΩ, 2000kΩ, 20000kΩ | ± 1.0% nke ịgụ + 2 mkpụrụ |
| Nnwale Diode | Ee | Gaa n'ihu voltage drop display |
| Nnwale na-aga n'ihu | Yes (with buzzer) | Buzzer sounds below approx. 50Ω |
Nkọwapụta izugbe
- Ogo nchekwa: CAT II 500V AC/DC, CAT III 300V
- Nchekwa ibu ibu: Fuses 500mA/500V (for VΩmA jack) and 10A/500V (for 10A jack)
- Ịnye ọkụ: 2 x AAA (LR03) batrị
- Ngosipụta: Dijitalụ LCD
- Akụkụ (L x W x H): Ihe dịka 11.8 x 6.5 x 3 cm (4.65 x 2.56 x 1.18 sentimita asatọ)
- Arọ: Approximately 100 grams (excluding batteries)
- Test Lead Length: Ihe dịka 75 cm
6. Nlekọta
6.1 Nhicha
Wipe the meter casing na adamp ákwà na ihe nhicha dị nro. Ejila ihe na-eme ka ihe dị nro ma ọ bụ ihe mgbaze dị nro. Hụ na mita ahụ kpọrọ nkụ kpamkpam tupu e jiri ya.
6.2 nnọchi batrị
When the low battery indicator appears on the display, replace the two AAA batteries as described in Section 3.1. Remove batteries if the meter is not used for an extended period to prevent leakage.
6.3 Mgbanwe Fuse
If the current measurement function stops working, the fuse may be blown. Fuses are located inside the meter. Fuse specifications are 500mA/500V for the VΩmA input and 10A/500V for the 10A input. Fuse replacement should only be performed by qualified personnel. Always replace with fuses of the exact same type and rating.
7. Nchọpụta nsogbu
- Enweghị Ngosipụta: Lelee ntinye batrị ma hụ na batrị anaghị agwụ agwụ.
- "OL" on Display: Indicates an overload or out-of-range measurement. Select a higher range or check for an open circuit (for continuity/resistance).
- Ọgụgụ na-ezighi ezi:
- Ensure test leads are correctly connected to the appropriate jacks.
- Verify the rotary switch is set to the correct function and range.
- Check if the batteries are low.
- Ensure the component or circuit is properly isolated for resistance/diode tests.
- Nha ugbu a anaghị arụ ọrụ: Check the fuse for the relevant current input (500mA or 10A).
8. Ozi nchekwa
This multimeter is designed to meet safety standards for electrical measuring instruments. However, improper use can result in electric shock or damage to the meter. Always follow these safety guidelines:
- Etinyekwala ihe karịrị ogo voltage between the input terminals or between any terminal and earth ground.
- Jiri nlezianya dị oke egwu mgbe ị na-arụ ọrụ na voltages karịrị 30V AC RMS, 42V elu, ma ọ bụ 60V DC. Vol ndị atages na-ebute ihe egwu ujo.
- Always disconnect the circuit power and discharge all high-voltage capacitors before measuring resistance or continuity.
- Ensure the rotary switch is in the correct position for the measurement being performed.
- Never measure current on a circuit with power applied if the test leads are connected to the "VΩmA" or "10A" jacks and the rotary switch is set to a voltage or resistance function. This can blow the fuse or damage the meter.
- Ejila mita ahụ ma ọ bụrụ na ọ dị ka ọ mebiri emebi ma ọ bụ ọ bụrụ na ụzọ ule mebiri emebi.
- Adhere to the CAT II 500V and CAT III 300V safety ratings. CAT III is for measurements in building installation (e.g., distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, e.g., stationary motors with permanent connection to the fixed installation). CAT II is for measurements performed on circuits directly connected to the low-voltage installation (e.g., household appliances, portable tools).
- Keep fingers behind the probe barriers during measurements.
9. Akwụkwọ ikike na nkwado
This Solight V15 II Multimeter is designed for reliability. For any warranty claims or technical support, please contact your retailer or the Solight customer service department. Please retain your proof of purchase for warranty purposes.





