Okwu mmalite
The AEMC Model 3711 is a portable clamp-on ground resistance tester designed for measuring electrical ground rod and grid resistance in multi-grounded systems. This instrument offers a non-intrusive method for testing without the need for auxiliary rods or disconnecting the ground system. It also measures leakage, neutral, and phase currents, providing a comprehensive solution for electrical system analysis.
Ozi nchekwa
Ịdọ aka ná ntị: Always read and understand all safety instructions before operating this instrument. Failure to do so may result in serious injury or death.
- This instrument is rated CAT III, indicating its suitability for measurements in building installations.
- Adhere to all local and national safety codes and regulations.
- Do not use the instrument if it appears damaged or is operating abnormally.
- Hụ na e mechiri ebe batrị ahụ nke ọma tupu e jiri ya.
- Avoid contact with live circuits during measurement.
- The instrument is double-insulated for enhanced safety.
Kedu ihe dị na Igbe ahụ
Chọpụta na ihe niile dị mgbe ị na-ebupụ:
- AEMC Ground Resistance Tester Model 3711
- Calibration Check Loop, 25Ω
- Akpa na-ebu siri ike
- 9V batrị alkaline
- Akwụkwọ ntuziaka (akwụkwọ a)
Ngwaahịa gafereview

Image: The AEMC Model 3711 Clamp-On Ground Resistance Tester. This image displays the device with its red clamp jaws open, a black body, and a green digital display showing "9.8 Ω". Control buttons are visible below the display.
The AEMC Model 3711 features an ergonomic design for one-handed operation. Its robust polycarbonate construction ensures durability for outdoor use. Key components include:
- Clamp Nkịta: 1.25" opening to accommodate ground rods and cables up to 1000kcmil.
- Ngosipụta LCD: Digital readout for resistance and current measurements.
- bọtịnụ njikwa: For power, data hold, and mode selection.
- Akụkụ batrị: Located at the rear for a 9V battery.
Ntọlite na ntinye batrị
- Open the battery compartment cover on the rear of the instrument.
- Insert the supplied 9V Alkaline battery, observing correct polarity.
- Mechie mkpuchi nke batrị batrị nke ọma.
- Press the "ON" button to power on the device. The display should illuminate.
Ntuziaka ọrụ
Ground Resistance Measurement
- Power on the AEMC 3711.
- Mepee clamp jaws by pressing the release lever.
- Debe ihe clamp around the ground conductor or rod to be tested, ensuring the jaws are fully closed.
- The instrument will automatically measure and display the ground resistance in Ohms (Ω).
- For stable readings, allow approximately one second for the measurement to settle.
- To hold the current reading on the display, press the "HOLD" button. Press again to release.
Ntụle ugbu a
- Power on the AEMC 3711.
- Mepee clamp agba.
- Debe ihe clamp around the conductor carrying the current to be measured (leakage, neutral, or phase current). Ensure the jaws are fully closed.
- The instrument will automatically measure and display the current in Amperes (A) ma ọ bụ milliAmpere (mA).
- The current measurement function provides direct readings for leakage current flowing to ground or circulating in ground loops (down to 1mA) and neutral/phase currents (up to 30A RMS).
Nlekọta na nhazigharị
Nhicha
Jiri nwayọ hichaa ngwa ahụ, damp akwa . Ejila ihe nchacha abrasive ma ọ bụ ihe mgbaze.
Ngbanwe batrị
When the low battery indicator appears on the display, replace the 9V battery as described in the "Setup and Battery Installation" section.
Nyochaa nhazi nhazi
Use the supplied 25Ω calibration check loop to verify the instrument's accuracy periodically. Clamp the tester around the loop; the reading should be approximately 25Ω. If the reading deviates significantly, contact AEMC customer support for service.
Nchekwa
Store the instrument in its hard carrying case in a dry, cool environment when not in use. Remove the battery if storing for extended periods.
Nchọpụta nsogbu
- No Display/Instrument Not Powering On: Lelee ntinye batrị ahụ ma hụ na batrị ahụ nwere ike ịchaji nke ọma. Dochie ya ma ọ bụrụ na ọ dị mkpa.
- Ọgụgụ na-adịghị akwụsi ike: Gbaa mbọ hụ na clamp jaws are fully closed around the conductor. Verify there is no excessive electrical noise in the environment.
- Ihe ngosi "OL" ma ọ bụ ibu oke ibu: The measured value exceeds the instrument's range. Ensure the correct measurement function is selected or the circuit parameters are within the device's limits.
- Ọgụgụ na-ezighi ezi: Perform a calibration check using the 25Ω loop. If the issue persists, contact technical support.
Nkọwapụta
| Oke | Uru |
|---|---|
| Ground Resistance Range | Autoranging 0.01 to 1200Ω |
| Oke nha ugbu a | Autoranging 1mA to 30.00A RMS |
| Resistance Measuring Frequency | 2403Hz |
| Current Measuring Frequency | 47 ruo 800Hz |
| Accuracy (Resistance) | ±2% reading ±0.02Ω (up to 25% of reading for 600-1200Ω) |
| Okpomọkụ na-arụ ọrụ | -10 ruo 55°C (14 ruo 131°F) |
| Isi Iyi Ike | Batrị 9V (gụnyere) |
| Ibu | 1kg (2.2lbs) |
| Akụkụ (H x W x D) | 235 x 100 x 55mm (9.25 x 3.94 x 2.17 sentimita) |
| Ọkụ na-emeghe | 1.25 sentimita asatọ |
| Ogo nchekwa | EN 61010, CAT III, UL approved, CE marked |
Akwụkwọ ikike na nkwado
For warranty information, technical support, or service inquiries, please refer to the official AEMC Instruments websaịtị ma ọ bụ kpọtụrụ ngalaba ọrụ ndị ahịa ha. Debe akwụkwọ nnata ịzụrụ maka akwụkwọ ikike.





