NATIONAL INSTRUMENTS PXI-6624 Counter Timer Module

Mgbakọ
- A na-eji nkwekọrịta ndị a na akwụkwọ a: Akara ngosi a na-egosi ndetu, nke na-eme ka ị mara ozi dị mkpa.
- italic: Ederede Italic na-egosi mgbanwe, nkwusi okwu ike, ntụaka obe, ma ọ bụ mmalite nke echiche isi. Ederede italic na-egosikwa ederede bụ ebe nchekwa maka okwu ma ọ bụ uru ị ga-eburịrị.
- monospace: Ederede dị na mkpụrụedemede a na-egosi ederede ma ọ bụ mkpụrụedemede ị kwesịrị itinye na ahụigodo, ngalaba nke koodu, mmemme examples, na syntax examples.
- A na-ejikwa font a maka aha draịva diski kwesịrị ekwesị, ụzọ, akwụkwọ ndekọ aha, mmemme, mmemme subprogram, subroutines, aha ngwaọrụ, ọrụ, arụmọrụ, mgbanwe, fileaha, na ndọtị.
Ngwa ngwa chọrọ
Calibrating the NI 6624 requires installing NI-DAQmx 7.5 or later on the calibration system. You can download NI-DAQmx at ni.com/downloads.NI-DAQmx supports programming an External Calibration in the LabVIEW, LabWindows™/CVI™, Microsoft Visual C++ 6.0, Microsoft VisualBasic 6.0, Microsoft .NET, and Borland C++ application development environments (ADEs). When you install NI-DAQmx, you only need to install support for the ADE that you intend to use.
Mara: NI na-atụ aro ka ị wụnye ngwa ọkwọ ụgbọala NI-DAQmx tupu ịwụnye ngwaọrụ NI 6624.
Achọrọ akwụkwọ
For information about NI-DAQmx and NI 6624, you can consult the following documents:
- NI-DAQmx Help—Nke a na-enyere aka file nwere ozi izugbe gbasara echiche nha, isi NI-DAQmx, yana ngwa nkịtị na-emetụta gburugburu mmemme niile.
- NI-DAQmx C Reference Help—Nke a na-enyere aka file contains C references and general information about measurement concepts.
- DAQ Getting Started guides—These guides describe how to installNI-DAQmx for Windows software and NI-DAQmx-supported DAQdevices, and how to confirm that your device is operating properly.
- NI 6624 User Manual—This document describes the electrical and mechanical aspects of the NI 6624 and contains information about its operation and programming.
- NI 6624 Specifications—This document lists the specifications for theNI 6624. The limits you use to verify the accuracy of the device are based on the specifications found in this document. You can download the latest version of this document from the NI Web saịtị na ni.com/manuals.
Eji NI-DAQmx tinye akwụkwọ ndị a. Ị nwekwara ike ịhụ ụdị akwụkwọ ọhụrụ kachasị ọhụrụ na ni.com/manuals.
Ogologo oge nhazi nhazi
NI na-atụ aro ka ị rụọ nhazi nhazi zuru oke ọbụlagodi otu ugboro kwa afọ. Ị nwere ike ibelata oge a dabere na mkpa ngwa gị ziri ezi.
Ngwa nnwale
NI na-atụ aro ka ị jiri akụrụngwa dị na Tebụl 1 maka ịhazi NI 6624.
Tebụl 1. Ngwa akwadoro
|
Akụrụngwa |
Ụdị akwadoro |
Ihe achọrọ kacha nta |
| Mpụga counter | PXI-6608 | £1 ppm uncertainty for measurement |
| +5 V Isi Iyi Ike | - | Must be able to source current of at least 1 A. |
| Discrete Resistors | - | 5% precision
Resistor values: 500 W (´ 2) |
| USB | SH100-100-F | - |
| ngọngọ njikọ | SCB-100 | - |
Mara: Ị ga-achọkwa ihe nkwụnye CompactPCI-to-PCI ma ọ bụrụ na ị na-emezi PXI-6624.
Soro ntuziaka ndị a ka ịkwalite akụrụngwa na gburugburu ebe obibi n'oge mmezi:
- Debe njikọ na ngwaọrụ dị mkpụmkpụ ka enwere ike. Ogologo eriri na waya na-arụ ọrụ dị ka antennae, na-ebuli mkpọtụ ọzọ nwere ike imetụta nha.
- Nyochaa na njikọ niile na ngwaọrụ ahụ, gụnyere njikọ ihu panel, nwere nchebe.
- Debe okpomọkụ nke gburugburu 25 Celsius. Okpomọkụ ngwaọrụ ga-adị ukwuu karịa okpomọkụ gburugburu.
- Debe iru mmiri dị n'okpuru 80%.
- Kwe ka oge ikpo ọkụ nke opekata mpe nkeji iri na ise iji hụ na sekit nke nha na-arụ ọrụ nke ọma.
- Use shielded copper wire for all cable connections to the device. Usetwisted-pairs wire to eliminate noise and thermal offsets.
- Gbaa mbọ hụ na edobere ọsọ fan chassis PXI ka ọ bụrụ HIGH, na ihe nzacha ndị fan dị ọcha, yana oghere ndị na-adịghị nwere oghere nwere oghere. Maka ozi ndị ọzọ, rụtụ aka na akwụkwọ ndetu mmanye-ikuku jụrụ oyi dị na ni.com/manuals.
- Tinye chassis na ngwa ahụ n'otu eriri ọkụ ka ịzena loops ala.
Pinout ngwaọrụ
Figure 1 shows the pinout of the NI 6624.
Usoro nhazi nhazi
When calibrating the NI 6624, you first use the 20 MHz timebase on theNI 6624 as the source to generate a 1 Hz square wave signal. The external counter measures the frequency of the square wave signal. You then calculate the frequency of the crystal oscillator to determine if the NI 6624 is operating within its specifications. Figure 2 illustrates the calibration process.
Ntụle ugboro ole nwere ejighị n'aka nke ruru nchikota nke igbu oge mgbasa nke NI 6624 na njehie nke counter mpụga. Ogologo oge mgbasa kachasị bụ 500 ns. N'iji oge nleta nke 1 s, enwere ike gbakọọ ejighị n'aka ịgbasa mgbasa dị ka ndị a:
500 ns/1 s = 0.5 ppm
The error of the external counter must be 1 ppm or less, as specified in the Test Equipment section of this document. The measurement uncertainty equals the sum of 0.5 ppm and 1 ppm, that is, 1.5 ppm. Therefore, the calculated frequency allows for a 1.5 ppm measurement uncertainty. Youcan reduce measurement uncertainty by using a more stable external counter, by increasing the measurement duration, or both.
Usoro nhazi nhazi gụnyere usoro ndị a:
- Initial Setup—Set up the test equipment for calibration.
- Verification—Verify the existing operation of the device. This step confirms whether the device is operating within its specifications.
Rụtụ aka na ntuziaka mmalite DAQ maka ozi gbasara otu esi etinye ngwanrọ na ngwaike yana otu esi ahazi ngwaọrụ na Measurement & Automation Explorer (MAX). NI na-atụ aro ka ị jiri PXI-6608 weta counter mpụga. Ị nwere ike iji counter ọzọ ma ọ bụrụ na ọ na-emezu ihe ndị a chọrọ ka akọwara na ngalaba Ngwa ule nke akwụkwọ a.
Mara: If you are calibrating the PXI-6624 on a PXI chassis, the measured frequency is that of the PXI backplane clock instead of the onboard crystal oscillator. To verify the onboard crystal oscillator, you must calibrate the PXI-6624 on a PCI chassis using a CompactPCI-to-PCI adapter
Mezue usoro ndị a ka ịtọlite akụrụngwa nnwale maka nhazigharị:
- USB NI 6624 gaa na ngọngọ njikọ 100-pin.
- Kpoo isi iyi ike +5 V na ngọngọ njikọ. Rụtụ aka na eserese 3 maka njikọ mmezi.
- a. Wụnye ọdụ ọkụ +5 V na ọdụ ngọngọ njikọ jikọtara na PFI 36 Vdd/CTR 0 Vdd pin (pin 7) na NI 6624.
- b. Waya ala nke isi iyi ike +5 V na ọdụ ngọngọ njikọ nke ejikọrọ na PFI 36 Vss/CTR 0 Vss pin (pin 8) na NI 6624.

- Wire a 500 Ω resistor between the connector block terminals that are connected to the PFI 36 Vdd/CTR 0 Vdd pin (pin 7) and PFI 36/CTR0 OUT pin (pin 9) on the NI 6624. Wire another 500 Ω resistor to the connector block terminal that is connected to the PFI 36/CTR 0 OUTpin (pin 9) on the NI 6624, as Figure 3 shows.
- Jikọọ counter mpụga na ngọngọ njikọ site na iji waya gbagọrọ agbagọ.
- a. Wire the input of the external counter to the other end of the 500 Ωresistor that is connected to the PFI 36/CTR 0 OUT pin (pin 9) onthe NI 6624.
- b. Wire the ground of the external counter to the connector block terminal that is connected to the PFI 36 Vss/CTR 0 Vss pin (pin 8)on the NI 6624.
Complete the following steps to verify the operation of the crystal oscillator the NI 6624:
- Configure counter 0 on the NI 6624 to generate a 1 Hz continuous square wave signal with a 50% duty cycle on the PFI 36/CTR 0 OUTpin (pin 9).
- Tulee ugboro nke mgbama emepụtara site na iji mpụta mpụta. Ọ nwere ike were oge maka counter mpụga iji tụọ mgbama 1 Hz.
- Jiri 20,000,000 mụbaa ugboro a tụrụ atụ iji gbakọọ ọnụọgụ mkpụrụ akụ kwesịrị ime n'otu sekọnd.
- Tụlee uru ugboro ole ị gbakọrọ site na nsonaazụ nha yana ụkpụrụ ndị a:
- a. If the frequency you measured falls between the upper limit of 20,001,000 Hz and the lower limit of 19,999,000 Hz, your device’s operating within its specifications.
- b. If the frequency you measured is above 20,001,000 Hz or below 19,999,000 Hz, your device is nonfunctional. Return the device toNI for repair or replacement.
Mara: Oke n'ime akwụkwọ a dabere na mbipụta Disemba 2006 nke nkọwapụta NI 6624. Rụtụ aka na nkọwapụta NI 6624 kacha ọhụrụ n'ịntanetị na ni.com/manuals. Ọ bụrụ na ụdị nkọwapụta dị nso nso a, gbakọọ oke dabere na nkọwapụta kachasị ọhụrụ.
Ebe aga maka nkwado
Ngwa ngwa mba Web site is your complete resource for technicalsupport. At ni.com/support you have access to everything from troubleshooting and application development self-help resources to emailand phone assistance from NI Application Engineers.National Instruments corporate headquarters is located at11500 North Mopac Expressway, Austin, Texas, 78759-3504.National Instruments also has offices located around the world to help address your support needs. For telephone support in the United States,create your service request at ni.com/support and follow the callinginstructions or dial 512 795 8248. For telephone support outside the UnitedStates, contact your local branch office:
- Australia 1800 300 800, Austria 43 662 457990-0,
- Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599,
- Canada 800 433 3488, China 86 21 5050 9800,
- Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,
- Finland 358 (0) 9 725 72511, France 01 57 66 24 24,
- Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737,
- Italy 39 02 41309277, Japan 0120-527196, Korea 82 02 3451 3400,
- Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710,
- Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466,
- New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60,
- Poland 48 22 328 90 10, Portugal 351 210 311 210,
- Russia 7 495 783 6851, Singapore 1800 226 5886,
- Slovenia 386 3 425 42 00, South Africa 27 0 11 805 8197,
- Spain 34 91 640 0085, Sweden 46 (0) 8 587 895 00,
- Switzerland 41 56 2005151, Taiwan 886 02 2377 2222,
- Thailand 662 278 6777, Turkey 90 212 279 3031,
- United Kingdom 44 (0) 1635 523545
CVI, Ngwa mba, NI, ni.com, na LabVIEW bụ ụghalaahịa nke National Instruments Corporation. Rụtụ aka na ngalaba Usoro ojiji na ni.com/legal. for more information about National Instruments trademarks. The mark LabWindows is used under a license from MicrosoftCorporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help» Patents in your software, the patent.txt. file na mgbasa ozi gị, ma ọ bụ National Instruments Patent Notice na ni.com/patent.
© 2009 National Instruments Corporation. Ikike niile echekwabara.
ỌRỤ MKWUTA: Anyị na-enye ọrụ nrụzi na nhazigharị asọmpi, yana akwụkwọ dị mfe ịnweta yana akụrụngwa enwere ike ibudata n'efu.
RERE MFỤỌTA GỊ
We buy new, used, decommissioned, and surplus parts from every Ni series. We work out the best solution to suit your individual needs
- Na-ere Maka Ego Ego
- Nweta kredit
- Nweta nkwekọrịta ahia ahia
ARSOLETE NI HARDWARE dị n'ahịa ma dị njikere ibupu: Anyị na-echekwa Ọhụrụ, Mgbakwunye Ọhụrụ, Nrụzigharị, yana Ngwa Niile emegharịrị
Rịọ nkwupụta PITA Ebe a (PXI-6624 Ntụle Ngwa Ngwa Mba/Modul Oge | Apex Waves) PXI-6624
Na-eme ka oghere dị n'etiti onye nrụpụta na sistemu ule ihe nketa gị.
Ụghalaahịa niile, ụdị na aha aha bụ ihe nke ndị nwe ha.
Akwụkwọ a nwere ozi maka ịhazi ngwaọrụ counter/ngụ oge nke mba 6624. Maka ozi ndị ọzọ gbasara nhazigharị, gaa na ni.com/calibration. ni.com/manuals.
Akwụkwọ / akụrụngwa
![]() |
NATIONAL INSTRUMENTS PXI-6624 Counter Timer Module [pdf] Ntuziaka onye ọrụ PXI-6624, PCI-6624, 6624, PXI-6624 Counter Timer Module, Counter Timer Module, Timer Module, Module |