HJA58

Richmat HJA58 Actuator Instruction Manual

Model: HJA58 | Brand: Generic (Manufactured by Richmat)

Okwu mmalite

This manual provides essential information for the installation, operation, and maintenance of the Richmat HJA58 Actuator. This actuator is designed as a replacement component for adjustable beds and hospital beds, specifically for the foot or head section. Please read this manual thoroughly before installation and use to ensure proper function and safety.

Ozi nchekwa

Ihe ngwugwu

Nyochaa na akụrụngwa niile dị na adịghị emebi emebi mgbe imepe ngwugwu ahụ.

Ntọala na nwụnye

Follow these steps carefully to replace your existing actuator. It is recommended to have a second person assist with lifting and positioning the bed frame if necessary.

  1. Prepare the Bed: Disconnect the adjustable bed from its power source. Ensure the bed is in a stable position, preferably flat or slightly elevated to access the actuator.
  2. Identify the Actuator: Locate the existing actuator on your adjustable bed. Confirm it is the HJA58 model by checking its label.
    Close-up of Richmat HJA58 actuator label showing model number, voltage, load, and duty cycle.

    Foto: nso nso view of the Richmat HJA58 actuator's identification label, displaying critical specifications such as item number HJA58, rated voltage DC 24V, max load 6000N, and duty cycle.

  3. Disconnect Old Actuator: Carefully disconnect the power cable and any mounting pins or bolts securing the old actuator to the bed frame. Note the orientation and connection points for reinstallation.
    Richmat HJA58 actuator with cable and connector visible.

    Image: The Richmat HJA58 actuator, showing the main body, extending rod, and the attached power cable with its connector, ready for installation or removal.

  4. Remove Old Actuator: Once disconnected, carefully remove the old actuator from the bed frame.
  5. Install New Actuator: Position the new Richmat HJA58 actuator in the same location and orientation as the old one. Secure it using the existing mounting pins or bolts. Ensure all connections are firm and stable.
    Richmat HJA58 foot actuator replacement for adjustable beds.

    Image: The Richmat HJA58 actuator, a linear motor designed for adjustable beds, shown in its compact form with the extending rod partially retracted.

  6. Njikọ Ike: Reconnect the actuator's power cable to the bed's control box or power supply. Ensure the connection is secure.
  7. Ọrụ nnwale: Reconnect the adjustable bed to its main power source. Carefully test the bed's functions (e.g., foot or head elevation) to ensure the new actuator operates smoothly and correctly.

Ntuziaka ọrụ

The Richmat HJA58 Actuator operates in conjunction with your adjustable bed's existing control system (remote control or integrated panel). Once properly installed and connected, the actuator will respond to commands from the bed's control unit to raise or lower the specific section of the bed it is attached to.

Nlekọta

The Richmat HJA58 Actuator requires minimal maintenance.

Nchọpụta nsogbu

If you encounter issues with your Richmat HJA58 Actuator, refer to the following common problems and solutions:

NsogbuIhe nwere ike imeNgwọta
Akụrụngwa anaghị agagharị.No power to the bed; loose cable connection; remote control issue; actuator overload/overheat.
  • Check if the bed is plugged into a working outlet.
  • Ensure all cables, especially the actuator's power cable, are securely connected.
  • Test the remote control (e.g., replace batteries if applicable).
  • Allow the actuator to cool down if it has been operated continuously (observe duty cycle).
Actuator moves slowly or with difficulty.Excessive load on the bed; mechanical obstruction; low voltage.
  • Reduce the load on the bed.
  • Check for any objects obstructing the bed's movement or the actuator's rod.
  • Ensure the power supply is providing adequate voltage.
Unusual noises from the actuator.Internal mechanical issue; loose mounting.
  • Ensure the actuator is securely mounted to the bed frame.
  • If noises persist and affect operation, discontinue use and contact support. Do not attempt to open the unit.

If the problem persists after attempting these solutions, please contact the seller or manufacturer for further assistance. Do not attempt to disassemble the actuator.

Nkọwapụta

NjirimaraNkọwa
NlereanyaHJA58
Onye nrụpụtaRichmat
NgwaAdjustable Beds, Hospital Beds (Foot or Head)
Ogo VoltageDC 24V
Max Ugbu a3A Max
Oke. Ibu ibuPush 6000N
Stroke/Installation Dimension70/245mm
Usoro ọrụ10%, Max. 2min ON / 18min OFF
Klas nchekwaIP20
Ụdị nkwụnyeIhe ndina Frame Mount
Ngwa gụnyere1 x mgbapụta
Nlekọta ngwaahịaJiri akwa akọrọ hichaa

Akwụkwọ ikike na nkwado

Specific warranty details for this product may vary depending on the point of purchase. Please refer to your purchase documentation or contact the seller directly for warranty information.

For technical support, questions regarding installation, or troubleshooting assistance, please reach out to the seller or manufacturer. When contacting support, please provide your product model (HJA58) and a detailed description of the issue.

Kpọtụrụ: Refer to your purchase receipt or the product listing for seller contact information.

Akwụkwọ ndị emetụtara - HJA58

Tupuview Richmat Bedding Product Overview: Actuators, Controllers, and Handsets
A keukwu overview of Richmat's bedding products, including linear and dual actuators, hand controls, controllers, and power supplies. Features detailed specifications and applications for enhancing comfort and functionality in bedding solutions.